HOMEExamination Instruments

L-7040 Sight Tester

L-4550 Internal Reading Lensmeter

K-0391 Universal Lightweight Trial Frame

We develop and manufacture various medical diagnostic equipments and supply them for domestic and overseas markets.
The sight tester (L-7040) can measure visual function and refractive errors in the most normal condition with bright and wide fields of vision.
A universal trial frame light weight type (K-0547) keeps patients in comfort during the long duration of the test.
Along with their basic functions, all of our examination instruments incorporate our original ideas for the sake of convenient use by opticians, medical doctors and patients.

L-7040 Sight Tester

SPEC

 

MEASUREMENT RANGE:

Sphere Measurement Range : '+16.75D to -19.00D two-sphere power dial combination

Increments : 0.25D steps

Cylinder Measurement Range : 0.00D to -6.00D(extendable to -8.00D):plus system optional

Increments : 0.25D steps(refinable to -0.125D steps)

Cylinder Axis Adjustment : 360°(double 0-180°scales)

Increments : 5°steps

PRISM VARIATION:

Rotary Prism Range : 20 Δ each(paired gives 40 Δ in any base direction)

Increments : 1 Δ scale graduation

LENSE:

Standard Cross-Cylinder : +/-0.25D

Optional Cross-Cylinder : +/-0.37D and +/-0.50D

AUXILIARY LENS DIALS:

0/0 : Open Aperture

R : Retinoscopic Lens +1.50D(+2.00D option)

P : Polarizing Filter

WMV : Maddox Rod,vertical:white(left).red(right)

WMH : Maddox Rod,horizontal:white(left).red(right)

GL : Green Lens(left)

RL : Red Lens(right)

+.12 : +.12D Spherical

PH : Pin Hole

10 Δ I : 10 Diopter Prism Base In(left)

6 Δ U : 6 Diopter Prism Base Up(right)

+/-.50 : +/-0.50D Fixed Cross Cylinder

OC : Occluder

DISTANCE:

Pupil Distance Adjustment : 48-75mm(1mm scale graduation)

Corneal Vertex Distance : 13.75mm

Reading Distance : 5-28"(plus cm &D scales)

DIMENSIONS AND WEIGHT:

Dimensions : 335(W)x97(D)x294(H)mm

Weight : 4.5kgs

Page Top

L-4550  Internal Reading Lensmeter

SPEC

 

COMPLETE ANALYSIS OF LENSES:

A.Determination of sphere power

Range of measurement of vertex power

0 to +/-25.00D.

Up to +/-3.00D. In 0.125D. steps and over +/-3.00D. In 0.25D. steps.

B.Determination of cylinder power

C.Determination of cylinder axis : 30°~180° in 1° step

D.Determination of prism power : 6 Δ (read in view field).

Prism compensator : Maximum 15 Δ in 1 Δ step.

INSTRUMENT TILTING ANGLE :

Freely adjustable between 30° ~ 90°

MEASURABLE LENS DIAMETER :

32mm ~ 90mm

LAMP :

110V,220V,240V-15W.(L4530V0)

DIMENSIONS AND WEIGHT:

Size : 160(W)x422(D)x335(H:at 30°)mm(D)
160(W)x422(D)x413(H:at 90°)mm(D)

Weight : 4.9kg

Page Top

K-0391 Universal Lightweight Trial Frame

SPEC

PD : 48 ~ 80mm (±0.8mm)

Cornea measuring plate : 2mm interval

Frame adjustable angles : 35°(upward 30°±5°, downward 5°+5°-2°)

Frame adjustable length : 35mm

Nose pad adjustable angles : 360°

Nose pad adjustable length : 22mm ±1.5mm

Maximum available lenses : 10 pcs

Axis of lens : 0°±5°, 180°±5°

Available colour : Black (K-0391), Violet (K-0391-V), White (K-0391-G)